Stability indicating RP-HPLC analytical method development and validation for the estimation of safinamide in bulk and marketed pharmaceutical dosage form
Abstract
A rapid and precise reverse phase high performance liquid chromatographic method has been developed for the validated of Safinamide, in its pure form as well as in tablet dosage form. Chromatography was carried out on a Symmetry C18 (4.6 x 150 mm, 5µm) column using a mixture of Methanol and water (45:55% v/v) as the mobile phase at a flow rate of 0.8ml/min, the detection was carried out at 260nm. The retention time of the Safinamide was 2.379±0.02 min respectively. The method produce linear responses in the concentration range of 24-120 mg/ml of Safinamide. The method precision for the determination of assay was below 2.0% RSD. The method is useful in the quality control of bulk and pharmaceutical formulations. The method was validated for accuracy, precision, linearity, robustness, ruggedness and LOD & LOQ of standard solution. The developed RP-HPLC method was found to be accurate, precise, linear, and robust and was successful applied to a pharmaceutical tablet formulation for qualitative estimation of Safinamide in Bulk form and Marketed Pharmaceutical Dosage forms.
How to Cite This Article
Dusakanti Akhila, Pasupuleti Sunitha, Vijaya Kuchana (2024). Stability indicating RP-HPLC analytical method development and validation for the estimation of safinamide in bulk and marketed pharmaceutical dosage form . International Journal of Multidisciplinary Research and Growth Evaluation (IJMRGE), 5(1), 93-99.